Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure

Use this identifier to reference this record.
Name:Description:Size:Format: 
RPC 2017 475.e1.pdf422.5 KBAdobe PDF Download

Advisor(s)

Abstract(s)

A 39-year-old woman underwent uneventful percutaneous occlusion of an ostium secundum atrial septal defect (ASD) with a 22 mm Ultrasept ASD Occluder®. Transesophageal echocardiography (TEE) performed two years after implantation revealed a de novo residual left-to-right shunt through the correctly implanted device. Three-dimensional transesophageal echocardiography (3D TEE) further clarified this finding by showing a perforation of the device membrane coating. The patient underwent transcatheter closure of the residual shunt with a 20 mm Ultrasept PFO®device. The procedure was guided by fluoroscopy and real-time 3D TEE. At the end of the procedure 3D TEE documented correct device deployment with complete defect coverage and absence of residual shunt.

Description

Keywords

HSM CAR HSM CAR PED Cardiac Surgical Procedures/methods Equipment Failure Heart Septal Defects, Atrial/surgery Remission Induction Septal Occluder Device

Citation

Rev Port Cardiol. 2017 Jun;36(6):475.e1-475.e3.

Research Projects

Organizational Units

Journal Issue

Publisher

Sociedade Portuguesa de Cardiologia

CC License

Altmetrics