dc.description.abstract | A 39-year-old woman underwent uneventful percutaneous occlusion of an ostium secundum atrial septal defect (ASD) with a 22 mm Ultrasept ASD Occluder®. Transesophageal echocardiography (TEE) performed two years after implantation revealed a de novo residual left-to-right shunt through the correctly implanted device. Three-dimensional transesophageal echocardiography (3D TEE) further clarified this finding by showing a perforation of the device membrane coating. The patient underwent transcatheter closure of the residual shunt with a 20 mm Ultrasept PFO®device. The procedure was guided by fluoroscopy and real-time 3D TEE. At the end of the procedure 3D TEE documented correct device deployment with complete defect coverage and absence of residual shunt. | pt_PT |
dc.description.version | info:eu-repo/semantics/publishedVersion | pt_PT |
dc.identifier.citation | Rev Port Cardiol. 2017 Jun;36(6):475.e1-475.e3. | pt_PT |
dc.identifier.doi | 10.1016/j.repc.2016.09.020 | pt_PT |
dc.identifier.uri | http://hdl.handle.net/10400.17/2935 | |
dc.language.iso | eng | pt_PT |
dc.peerreviewed | yes | pt_PT |
dc.publisher | Sociedade Portuguesa de Cardiologia | pt_PT |
dc.subject | HSM CAR | pt_PT |
dc.subject | HSM CAR PED | pt_PT |
dc.subject | Cardiac Surgical Procedures/methods | pt_PT |
dc.subject | Equipment Failure | pt_PT |
dc.subject | Heart Septal Defects, Atrial/surgery | pt_PT |
dc.subject | Remission Induction | pt_PT |
dc.subject | Septal Occluder Device | |
dc.title | Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure | pt_PT |
dc.title.alternative | Encerramento Percutâneo de Comunicação Interauricular Residual por Falência de Dispositivo | pt_PT |
dc.type | journal article | |
dspace.entity.type | Publication | |
oaire.citation.endPage | 475.e3 | pt_PT |
oaire.citation.issue | 6 | pt_PT |
oaire.citation.startPage | 475.e1-475.e3 | pt_PT |
oaire.citation.title | Revista Portuguesa de Cardiologia | pt_PT |
oaire.citation.volume | 36 | pt_PT |
rcaap.rights | openAccess | pt_PT |
rcaap.type | article | pt_PT |